Services Available |
---|
The CSAT3 is a three-dimensional (3D) sonic anemometer that measures three orthogonal wind components and the speed of sound. In eddy covariance systems, it can measure the turbulent fluctuations of horizontal and vertical wind. These measurements are then used to calculate momentum flux and friction velocity. This sonic anemometer can also provide average horizontal wind speed and direction measurements. The CSAT3 comes with 7.6 M cables.
The CSAT3 has a 10 cm vertical measurement path and operates in a pulsed acoustic mode. The three orthogonal wind components (ux, uy, uz) and the speed of sound (c) are measured and output at a maximum rate of 60 Hz. Analog outputs and two types of digital outputs are provided.
Measurements can be triggered from from the CSAT3’s internal clock, the PC-generated RS-232 command, or the datalogger’s SDM command. The SDM protocol supports a group trigger for synchronizing multiple CSAT3s.
The FW05 fine wire thermocouple (12.7 μm diameter) is available as an option for fast response temperature measurements.
Measurement Path Length |
|
Path Angle from Horizontal | 60° |
Construction | Sealed sonic transducers and electronics |
Anemometer Head Materials | Stainless-steel tubing |
Electronics Box Materials | Cast aluminum |
Operating Temperature Range | -30° to +50°C |
Voltage Supply | 10 to 16 Vdc |
Current |
|
Digital SDM Output Signal | CSI 33.3 k baud serial interface for data logger/sensor communication. (Data type is 2-byte integer per output plus 2-byte diagnostic.) |
Support Arm Diameter | 1.59 cm (0.63 in.) |
Transducer Diameter | 0.64 cm (0.25 in.) |
Transducer Mounting Fingers Diameter | 0.84 cm (0.33 in.) |
Cable Length | 7.62 m (25 ft) |
Anemometer Head Dimensions | 47.3 x 42.4 cm (18.6 x 16.7 in.) |
Anemometer Head Weight | 1.7 kg (3.7 lb) |
Electronics Box Dimensions | 26 x 16 x 9 cm (10.24 x 6.3 x 3.54 in.) |
Electronics Box Weight | 3.8 kg (8.4 lb) |
Measurements |
|
Outputs |
ux, uy, uz, c (ux, uy, uz are wind components referenced to the anemometer axes; c is speed of sound.) |
Speed of Sound | Determined from three acoustic paths; corrected for crosswind effects. |
Measurement Rate | Programmable from 1 to 60 Hz, instantaneous measurements. Two over-sampled modes are block averaged to either 20 Hz or 10 Hz. |
Measurement Resolution |
|
Offset Error |
|
Gain Error |
|
Wind Direction Accuracy | ±0.7° at 1 m/s (for horizontal wind) |
Rain | Innovative ultrasonic signal processing and user-installable wicks considerably improve the performance of the anemometer under all rain events. |
Digital RS-232 Output Signal |
|
Baud Rate | 9600, 19200 bps |
Data Type | 2-byte integer per output plus 2-byte diagnostic |
Analog |
|
Number of Outputs | 4 |
Voltage Range | ±5 V |
Number of Bits | 12 |
SDM & RS-232 Digital Outputs Reporting Range |
|
Full-Scale Wind | ±65.535 m/s autoranging between four ranges (Least significant bit is 0.25 to 2 mm/s.) |
Speed of Sound | 300 to 366 m/s (-50° to +60°C) Least significant bit is 1 mm/s (0.002°C). |
Analog Outputs Reporting Range |
|
ux | ±30 m s-1, ±60 m s-1 |
uy | ±30 m s-1, ±60 m s-1 |
uz | ±8 m s-1 |
c | 300 to 366 m s-1 (-50 to +60°C) |
Analog Outputs LSB |
|
ux | 15 mm s-1, 30 mm s-1 |
uy | 15 mm s-1, 30 mm s-1 |
uz | ±8 m s-1 |
c | 16 mm s-1 (0.026°C) |
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
Product | Compatible | Note |
---|---|---|
21X (retired) | ||
CR10 (retired) | ||
CR1000 (retired) | ||
CR10X (retired) | ||
CR200X (retired) | ||
CR211X (retired) | ||
CR216X (retired) | ||
CR23X (retired) | ||
CR300 | ||
CR3000 (retired) | ||
CR500 (retired) | ||
CR5000 (retired) | ||
CR510 (retired) | ||
CR6 | ||
CR800 (retired) | ||
CR850 (retired) | ||
CR9000 (retired) | ||
CR9000X (retired) |
Data Logger Model | Data Logger Operating System (OS) or PROMs Required |
CSAT3 OS Required |
21X | 1K 6145-3, 6146-5, 10006-1 | 2.1 or higher |
21X | 2K 6148-5, 6149-7, 6070-64 | 2.1 or higher |
CR800, CR850 | All OS | 3.0 or higher |
CR10 | 2K 5954-396; 4K is a Library Special |
2.1 or higher |
CR10X | OS 1.4 or higher | 2.1 or higher |
CR1000 | All OS | 3.0 or higher |
CR23X | All OS | 2.1 or higher |
CR3000 | All OS | 3.0 or higher |
CR5000 | All OS | 3.0 or higher |
CR9000 | 2.01 or higher | 2.1 or higher |
CR9000X | All OS | 2.1 or higher |
Number of FAQs related to CSAT3: 4
Expand AllCollapse All
The sensors used in the eddy-covariance application are not compatible with SCWin. An open-path eddy-covariance program is available for purchase as pn 18442, CRBasic Basic Eddy Covariance Program. This is the program without energy balance sensors. Also available is pn 18443, CRBasic Extended Eddy Covariance Program, with energy balance sensors. To order a custom configured program, contact the Environmental Group for assistance.
Yes. The bubble level can be replaced in the field using pn 31962, 1-Ring Replacement Bubble Level with Mounting Bumper and Hardware.
With long cable lengths, monitor the SDM data collected from the CSAT3. If the CSAT3 diagnostic word is 61441 or NAN, there are SDM communications errors. If the SDM bit period is increased, SDM communications should be restored. Use the SDMSpeed() instruction to change the SDM bit period. Use the following values for the listed data loggers:
The CSAT3/3A/3H is compatible with three output signals: analog, RS-232, and Synchronous Device for Measurement (SDM). The sonic anemometer can be interfaced to any data-acquisition system that is compatible with analog measurements or RS-232 serial communications. SDM is used with Campbell Scientific data loggers.
We've updated our privacy policy. Learn More