The 247 was similar to the 247W-L, except it did not have a stainless steel sleeve, and therefore had a smaller diameter. The 247W measured conductivity using three passivated 316 stainless steel annular electrodes with dc isolation capacitors. A thermistor measured temperature.
The "-L" on a product model indicates that the cable length is specified at the time of order.
The 247 had a pH range of 3.0 to 9.0, with a conductivity accuracy of ±5% (at 25°C) over 0.45 to 7.0 mS/cm range, ±10% (at 25°C) over 0.005 to 0.45 mS/cm range. Temperature polynomial linearization accuracy was typically less than ±0.1°C over 0° to +48°C range. Interchangeability accuracy was typically less than ±0.2°C over 0° to +50°C range.
The 247-L (pn 7343) was originally replaced with the CS547, which was replaced with the CS547A.
Operating Temperature | 0° to 50°C |
Diameter | 1.91 cm (0.75 in.) |
Length | 7.938 cm (3.125 in.) |
Weight | 0.2 kg (0.5 lb) with 3.05-m (10-ft) cable |
The 247 required one differential analog input, one single-ended analog input, and two excitation channels.